BX41M/BX51M金相显微镜BX41M/BX51M
|
|
|
|
公司名称:
东莞市日昶计量仪器有限公司
|
|
参考价格:
80000
元
|
|
发布时间:
2016-12-21
|
| 产品描述: |
| BX41M BX41M-ESD
Optical system UIS optical system
Objectives UIS objectives
Eyepieces UIS eyepieces
Microscope frame Illuminator Reflected light Reflected light(with ESD performance)
6V30W power supply, light preset switch 6V30W power supply, light preset switch
Focus Vertical stage movement: 25mmFine focus movement: 100mMinimum graduation: 1mUpper limit stopTorque adjustment for coarse adjustment knobs
Maximum specimen height 65mm(w/o spacer)
Observation tube Widefield(F.N. 22) Inverted: binocular, trnocular and tilting binocular observation tubesErect: trinocular and tilting binocular observation tubes
Super widefield(F.N. 26.5) Inverted: trinocular observation tubeErect: trinocular and tilting trinocular observation tubes
Reflected light illuminator BX-KMA BX-KMA-ESD
30W halogen, BF/DIC/KPO 30W halogen, BF/DIC/KPO
Revolving nosepiece For BF Sextuple, centerable sextuple, septuple revolving nosepieces(optional: motorized sextuple) Sextuple revolving nosepiece with slider slot for DIC with ED performance
Stage Coaxial stage with left or right hand low drive control, with torque adjustment(Movement: X=76mm, Y=52mm)Large-sized coaxial stage with left or right hand low drive control,With Y direction locking mechanism)Movement: X=110, Y=52mm)
附件
Eyepiece Magnification 10X, erect image (inverted when used with upright observation tube), F. N. 14. Diopter adjustment range: ±5 diopter. Provided with rubber eye shade.
Measuring scale Scale lines graduated in increments of 1mm in the entire 10mm length. Shift of scale lines: 1mm per rotation of the shift ring, the circumference of which is divided into 100 graduations.
Measuring range 10mm/objective magnification
Compensation limit for objective magnification tolerance ±5% by combined use of the zoom compensation ring and the provided stage micrometer. Compensation ring clamping screw. Magnification compensation scale.
Actual size Actual size(mm) = measured value(mm)/objective magnification
Reproducibility Reproducibility error ±0.007/A mm(A…objective magnification)
Accuracy Measuring error(A…objective magnification, L…measured length in mm)±[ (0.0002XA + 0.002)L + 0.007/A] mm
10/100 10mm in 100divisions
Cross 10/100 100 in 100 divisions on crosslines
H5/5 5mm in 5 divisions in grid pattern
H7/7 7mm in 7 divisions in grid pattern
H10/100 10mm in 100 divisions in grid pattern
|
|
|
|
| |
| |
BX41M/BX51M金相显微镜BX41M/BX51M的详细介绍
| |
BX41M |
BX41M-ESD |
| Optical system |
|
UIS optical system |
| Objectives |
UIS objectives |
| Eyepieces |
UIS eyepieces |
| Microscope frame |
Illuminator |
Reflected light |
Reflected light(with ESD performance) |
| 6V30W power supply, light preset switch |
6V30W power supply, light preset switch |
| |
Focus |
Vertical stage movement: 25mmFine focus movement: 100mMinimum graduation: 1mUpper limit stopTorque adjustment for coarse adjustment knobs |
| |
Maximum specimen height |
65mm(w/o spacer) |
| Observation tube |
Widefield(F.N. 22) |
Inverted: binocular, trnocular and tilting binocular observation tubesErect: trinocular and tilting binocular observation tubes |
| |
Super widefield(F.N. 26.5) |
Inverted: trinocular observation tubeErect: trinocular and tilting trinocular observation tubes |
| Reflected light illuminator |
BX-KMA |
BX-KMA-ESD |
| 30W halogen, BF/DIC/KPO |
30W halogen, BF/DIC/KPO |
| Revolving nosepiece |
For BF |
Sextuple, centerable sextuple, septuple revolving nosepieces(optional: motorized sextuple) |
Sextuple revolving nosepiece with slider slot for DIC with ED performance |
| Stage |
Coaxial stage with left or right hand low drive control, with torque adjustment(Movement: X=76mm, Y=52mm)Large-sized coaxial stage with left or right hand low drive control,With Y direction locking mechanism)Movement: X=110, Y=52mm) |
| 附件 |
|

| Eyepiece |
Magnification 10X, erect image (inverted when used with upright observation tube), F. N. 14. Diopter adjustment range: ±5 diopter. Provided with rubber eye shade. |
| Measuring scale |
Scale lines graduated in increments of 1mm in the entire 10mm length. Shift of scale lines: 1mm per rotation of the shift ring, the circumference of which is divided into 100 graduations. |
| Measuring range |
10mm/objective magnification |
| Compensation limit for objective magnification tolerance |
±5% by combined use of the zoom compensation ring and the provided stage micrometer. Compensation ring clamping screw. Magnification compensation scale. |
| Actual size |
Actual size(mm) = measured value(mm)/objective magnification |
| Reproducibility |
Reproducibility error ±0.007/A mm(A…objective magnification) |
| Accuracy |
Measuring error(A…objective magnification, L…measured length in mm)±[ (0.0002XA + 0.002)L + 0.007/A] mm |
| 10/100 |
10mm in 100divisions |
| Cross 10/100 |
100 in 100 divisions on crosslines |
| H5/5 |
5mm in 5 divisions in grid pattern |
| H7/7 |
7mm in 7 divisions in grid pattern |
| H10/100 |
10mm in 100 divisions in grid pattern |
|
|
|
|
|