G E N T E C - E O
ENERGY DETECTORS
General Use
• Extremely high damage threshold: 2-4 J/cm2
• Durable
• Versatile
• Full NIST-traceability
• Smart Interface
ED-200+
This is the most versatile energy detector
in the Gentec-EO family of products. The
new specially formulated coating of the
ED-200+ Series provides one of the most
damage resistant energy detectors in
the world. It provides the best combination
of damage resistance, broad
bandwidth and sensitivity that is available
anywhere. The high average power density
and very high pulse energy density make
this the preferred choice for concentrated
high energy beams.
You can read them with Gentec-EO SOLO
and DUO monitors as well as with any
oscilloscope. They are also compatible
with the EDX-1 amplifier. The amplifier
doubles the repetition rate capability of
these detectors as well as providing a
variable gain from 1 to 1000.
EDA-200 Attenuator
The EDA-200 was designed to extend the
performance of the ED-200+ energy
detectors in the spectral range of 400 nm
to 2.5 μm. Nominally, 50% of the incident
energy is reflected, 30 % is absorbed and
20 % transmitted to the energy detector.
It handles a peak pulse energy density
of at least 85 J/cm2 for long pulse
(150μs) and 7 J/cm2 for short pulses
(7ns). Designed for easy installation it
simply snaps onto the ED-200+.
ED-200L+
We put the same tough, broadband, and
sensitive absorber in this low-profile
package to make a detector especially
suited for installation in tight optical
setups. The combination of features
provided by this versatile detector
makes it right for a wide range of OEM
applications.
ED-200LA+
To measure higher energies, handle
higher average power beams and still
keep the streamlined profile, we have
combined the ED-200L+ with an
improved attenuator good to 7 J/cm2.
The attenuator is in a mount made
specifically for the thin profile detector.
ED-200LA+ ED-200+
ED-200+ ED-200L+ ED-200LA+
MEASUREMENT CAPABILITY
Spectral Range 0.19 – 11 μm 0.19 – 11 μm 0.19 – 2.5 μm
Maximum Measurable Energy
7 ns pulse, 1.064 μm a 7.6 J 7.6 J 30 J
Minimum Measurable Energy b 200 μJ 200 μJ 1 mJ
Noise Equivalent Energy 10 μJ 10 μJ 50 μJ
Sensitivity (typical) c,d 10 V/J 10 V/J 2 V/J
Max Repetition Frequency 100 Hz / 200 Hz e 100 Hz / 200 Hz e 100 Hz / 200 Hz e
Maximum Pulse Widthf 0.8 – 5 ms 0.8 – 5 ms 0.8 – 5 ms
Rise Time (typical) 1.5 ms 1.5 ms 1.5 ms
Calibration Uncertainty g ± 5 % ± 5 % ± 5 %
Repeatability (precision) < 0.5 % < 0.5 % < 0.5 %
DAMAGE THRESHOLDS
Max Average Power (continuous) 10 W 5 W 20 W
Max Average Power Density 2 kW/cm2 2 kW/cm2 10 kW/cm2
Max Energy Density
10.6 μm, 7 ns, 10 Hz 4 J/cm2 4 J/cm2
1.064 μm, 150 μs, 10 Hz 85 J/cm2
1.064 μm, 7 ns, 10 Hz 2 J/cm2 2 J/cm2 7 J/cm2
532 nm, 7 ns, 10 Hz 2 J/cm2 2 J/cm2 6 J/cm2
266 nm, 7 ns, 10 Hz 160 mJ/cm2 160 mJ/cm2
PHYSICAL CHARACTERISTICS
Effective Aperture Diameter 22 mm Ø 22 mm Ø 22 mm Ø
Absorber High Energy High Energy High Energy + Attenuator
Dimensions 74.8 x Ø43.7 mm 14 x Ø55 mm 19 x Ø55 mm
Weight (head only) 240 g 95 g 120 g
Effective Area 3.8 cm2 3.8 cm2 3.8 cm2
Specifications subject to change without notice.
PRINTED IN CANADA
ED-200+, ED-200L+ AND ED-200LA+ SPECIFICATIONS
ED-200+ ED-200L+ ED-200LA+
• CO2
• Er:YAG
• Ho:YAG
• Ruby
• Small Excimer
• CO2
• Er:YAG
• Ho:YAG
• Nd:Glass
• Small Excimer
• Nd:YAG
• Ho:YAG
• Tm:YAG
• YLF
• Ruby (long pulse)
• Opthamology
• Angioplasty
• Drilling
• OEM
• Cardiac surgery
• Robotics
• Micromachining
• Urology
• Dermatology
a. Increasing the pulse width increases the maximum measurable energy.
b. Nominal value, actual value depends on electrical noise in the measurement system.
c. Load: 1 MΩ and ≤130 pF |
|
|
|
|
|
PT1-ED200能量探头的详细介绍
G E N T E C - E O
ENERGY DETECTORS
General Use
• Extremely high damage threshold: 2-4 J/cm2
• Durable
• Versatile
• Full NIST-traceability
• Smart Interface
ED-200+
This is the most versatile energy detector
in the Gentec-EO family of products. The
new specially formulated coating of the
ED-200+ Series provides one of the most
damage resistant energy detectors in
the world. It provides the best combination
of damage resistance, broad
bandwidth and sensitivity that is available
anywhere. The high average power density
and very high pulse energy density make
this the preferred choice for concentrated
high energy beams.
You can read them with Gentec-EO SOLO
and DUO monitors as well as with any
oscilloscope. They are also compatible
with the EDX-1 amplifier. The amplifier
doubles the repetition rate capability of
these detectors as well as providing a
variable gain from 1 to 1000.
EDA-200 Attenuator
The EDA-200 was designed to extend the
performance of the ED-200+ energy
detectors in the spectral range of 400 nm
to 2.5 μm. Nominally, 50% of the incident
energy is reflected, 30 % is absorbed and
20 % transmitted to the energy detector.
It handles a peak pulse energy density
of at least 85 J/cm2 for long pulse
(150μs) and 7 J/cm2 for short pulses
(7ns). Designed for easy installation it
simply snaps onto the ED-200+.
ED-200L+
We put the same tough, broadband, and
sensitive absorber in this low-profile
package to make a detector especially
suited for installation in tight optical
setups. The combination of features
provided by this versatile detector
makes it right for a wide range of OEM
applications.
ED-200LA+
To measure higher energies, handle
higher average power beams and still
keep the streamlined profile, we have
combined the ED-200L+ with an
improved attenuator good to 7 J/cm2.
The attenuator is in a mount made
specifically for the thin profile detector.
ED-200LA+ ED-200+
ED-200+ ED-200L+ ED-200LA+
MEASUREMENT CAPABILITY
Spectral Range 0.19 – 11 μm 0.19 – 11 μm 0.19 – 2.5 μm
Maximum Measurable Energy
7 ns pulse, 1.064 μm a 7.6 J 7.6 J 30 J
Minimum Measurable Energy b 200 μJ 200 μJ 1 mJ
Noise Equivalent Energy 10 μJ 10 μJ 50 μJ
Sensitivity (typical) c,
|
|
|
|